Title |
Diagnostic accuracy of low and high tube voltage coronary CT angiography using an X-ray tube potential-tailored contrast medium injection protocol
|
---|---|
Published in |
European Radiology, November 2017
|
DOI | 10.1007/s00330-017-5150-z |
Pubmed ID | |
Authors |
Moritz H. Albrecht, John W. Nance, U. Joseph Schoepf, Brian E. Jacobs, Richard R. Bayer, Sheldon E. Litwin, Michael A. Reynolds, Katharina Otani, Stefanie Mangold, Akos Varga-Szemes, Domenico De Santis, Marwen Eid, Georg Apfaltrer, Christian Tesche, Markus Goeller, Thomas J. Vogl, Carlo N. De Cecco |
Abstract |
To compare the diagnostic accuracy between low-kilovolt peak (kVp) (≤ 100) and high-kVp (> 100) third-generation dual-source coronary CT angiography (CCTA) using a kVp-tailored contrast media injection protocol. One hundred twenty patients (mean age = 62.6 years, BMI = 29.0 kg/m2) who underwent catheter angiography and CCTA with automated kVp selection were separated into two cohorts (each n = 60, mean kVp = 84 and 117). Contrast media dose was tailored to the kVp level: 70 = 40 ml, 80 = 50 ml, 90 = 60 ml, 100 = 70 ml, 110 = 80 ml, and 120 = 90 ml. Contrast-to-noise ratio (CNR) was measured. Two observers evaluated image quality and the presence of significant coronary stenosis (> 50% luminal narrowing). Diagnostic accuracy (sensitivity/specificity) with ≤ 100 vs. > 100 kVp CCTA was comparable: per patient = 93.9/92.6% vs. 90.9/92.6%, per vessel = 91.5/97.8% vs. 94.0/96.8%, and per segment = 90.0/96.7% vs. 90.7/95.2% (all P > 0.64). CNR was similar (P > 0.18) in the low-kVp vs. high-kVp group (12.0 vs. 11.1), as ws subjective image quality (P = 0.38). Contrast media requirements were reduced by 38.1% in the low- vs. high-kVp cohort (53.6 vs. 86.6 ml, P < 0.001) and radiation dose by 59.6% (4.3 vs. 10.6 mSv, P < 0.001). Automated tube voltage selection with a tailored contrast media injection protocol allows CCTA to be performed at ≤ 100 kVp with substantial dose reductions and equivalent diagnostic accuracy for coronary stenosis detection compared to acquisitions at > 100 kVp. • Low-kVp coronary CT angiography (CCTA) enables reduced contrast and radiation dose. • Diagnostic accuracy is comparable between ≤ 100 and > 100 kVp CCTA. • Image quality is similar for low- and high-kVp CCTA. • Low-kVp image acquisition is facilitated by automated tube voltage selection. • Tailoring contrast injection protocols to the automatically selected kVp-level is feasible. |
X Demographics
Geographical breakdown
Country | Count | As % |
---|---|---|
Unknown | 1 | 100% |
Demographic breakdown
Type | Count | As % |
---|---|---|
Members of the public | 1 | 100% |
Mendeley readers
Geographical breakdown
Country | Count | As % |
---|---|---|
Unknown | 21 | 100% |
Demographic breakdown
Readers by professional status | Count | As % |
---|---|---|
Other | 3 | 14% |
Student > Bachelor | 3 | 14% |
Student > Ph. D. Student | 3 | 14% |
Researcher | 2 | 10% |
Professor | 1 | 5% |
Other | 2 | 10% |
Unknown | 7 | 33% |
Readers by discipline | Count | As % |
---|---|---|
Medicine and Dentistry | 6 | 29% |
Biochemistry, Genetics and Molecular Biology | 2 | 10% |
Nursing and Health Professions | 2 | 10% |
Social Sciences | 1 | 5% |
Psychology | 1 | 5% |
Other | 2 | 10% |
Unknown | 7 | 33% |