The success of needle decompression for tension pneumothorax is variable, and there are no objective measures assessing effective decompression. Colorimetric capnography, which detects carbon dioxide present within the pleural space, may serve as a simple test to assess effective needle decompression.
Three swine underwent traumatically induced tension pneumothorax, (n=15 standard of care, n=15 with needle capnography). Standard of care needle thoracostomy (n=15) using an 8 cm angiocatheter was performed. Similarly, decompression was performed with the addition of colorimetric capnography. Subjective operator assessment of decompression was recorded and compared to true decompression using thoracoscopic visualization for both techniques. Area under receiver operating curves was calculated and pairwise comparison was used to assess statistical significance (p<0.05).
Needle colorimetric capnography's detection of decompression was found to be 100% accurate (15/15 attempts), when compared to that of thoracoscopic assessment (true decompression). Furthermore, it accurately detected the lack of tension pneumothorax, no presence of pathologic/space occupying lesion, in 100% (10/10 attempts) of cases. Standard of care needle decompression was detected by operators 9/15 attempts (60%) and detected when tension pneumothorax not present 3/10 attempts (30%). True decompression occurred using colorimetry 15/15 (100%) times and using standard of care treatment 12/15 (80%) times. Area under receiver operating curves were 0.65 for standard of care and 1.0 for needle capnography (p=0.002) CONCLUSION: Needle decompression with colorimetric capnography provides a rapid, effective, and highly accurate method for eliminating operator bias for tension pneumothorax decompression. This may have utility for treatment of this life threatening condition.