Hypophosphatemia is commonly observed in critically ill patients. Inorganic phosphorus is quantified by spectrophotometric measurement of a phosphomolybdate complex, a method with multiple documented interferents. Our clinical laboratory was contacted to investigate a case of asymptomatic hypophosphatemia in a patient receiving high-dose liposomal amphotericin B therapy (L-AMB).
In vitro experiments were performed by spiking L-AMB into residual plasma specimens. Phosphate was measured on the Beckman Coulter AU and Ortho Diagnostics Vitros instruments.
When measured on the AU, phosphate in plasma with approximately 250mcg/mL of L-AMB demonstrated a median negative bias of 3.45mg/dL relative to unspiked samples. In contrast, Vitros phosphate measurements demonstrated excellent agreement for specimens with and without L-AMB (median bias -0.2mg/dL).
High L-AMB concentrations induced a significant negative bias on phosphate measured by the AU assay, but did not affect the Vitros assay. Laboratorians and clinicians should be aware of this phenomenon in patients receiving L-AMB who develop unexplained hypophosphatemia.