To understand the effective radiation dose during percutaneous cryoablation (CA) and radiofrequency ablation (RFA) and characterize variables which may affect individual dose.
Effective radiation dose was determined by conversion of the dose-length product (DLP) from CT scans performed during percutaneous cryoablation (CA) or radiofrequency ablation (RFA) for patients with solitary renal masses (<4cm) at 4 academic centers. Radiation dose per case was compared between patients and institutions using multivariate and univariate analysis. Lifetime attributable risk (LAR) of cancer was calculated for each institution and utilized to determine the number needed to harm (NNH) for a range of ages at the time of exposure.
One hundred twenty-three patients met inclusion criteria with a mean age of 71 years. Sixty-nine percent of patients were male, mean BMI was 29.4, and mean tumor size was 2.2 cm. The mean effective radiation dose per ablation was 40 mSv (range 3.7-147). On multivariate analysis only BMI and institution were associated with radiation dose. No significant difference in radiation exposure was seen for RFA or CA procedures.
Radiation exposure during percutaneous ablation is similar to a multi-phase CT scan. However, there is wide variability in individual treatment exposure, varying from 3.7-147 mSv, depending primarily on institution and BMI. Standardization of protocols is required in order to achieve as low as reasonably achievable (ALARA) levels of radiation.