↓ Skip to main content

Novel control system to prevent right ventricular failure induced by rotary blood pump

Overview of attention for article published in Journal of Artificial Organs, February 2014
Altmetric Badge

Readers on

mendeley
24 Mendeley
You are seeing a free-to-access but limited selection of the activity Altmetric has collected about this research output. Click here to find out more.
Title
Novel control system to prevent right ventricular failure induced by rotary blood pump
Published in
Journal of Artificial Organs, February 2014
DOI 10.1007/s10047-014-0757-1
Pubmed ID
Authors

Mamoru Arakawa, Takashi Nishimura, Yoshiaki Takewa, Akihide Umeki, Masahiko Ando, Yuichiro Kishimoto, Yutaka Fujii, Shunei Kyo, Hideo Adachi, Eisuke Tatsumi

Abstract

Right ventricular (RV) failure is a potentially fatal complication after treatment with a left ventricular assist device (LVAD). Ventricular septal shift caused by such devices is an important factor in the progress of RV dysfunction. We developed a control system for a rotary blood pump that can change rotational speed (RS) in synchrony with the cardiac cycle. We postulated that decreasing systolic RS using this system would alter ventricular septal movement and thus prevent RV failure. We implanted the EVAHEART ventricular assist device into seven adult goats weighing 54.1 ± 2.1 kg and induced acute bi-ventricular dysfunction by coronary embolization. Left and RV pressure was monitored, and ventricular septal movement was echocardiographically determined. We evaluated circuit-clamp mode as the control condition, as well as continuous and counter-pulse modes, both with full bypass. As a result, a leftward ventricular septal shift occurred in continuous and counter-pulse modes. The septal shift was corrected as a result of decreased RS during the systolic phase in counter-pulse mode. RV fractional area change improved in counter-pulse (59.0 ± 4.6%) compared with continuous (44.7 ± 4.0%) mode. In conclusion, decreased RS delivered during the systolic phase using the counter-pulse mode of our new system holds promise for the clinical correction of ventricular septal shift resulting from a LVAD and might confer a benefit upon RV function.

Mendeley readers

Mendeley readers

The data shown below were compiled from readership statistics for 24 Mendeley readers of this research output. Click here to see the associated Mendeley record.

Geographical breakdown

Country Count As %
United States 1 4%
Unknown 23 96%

Demographic breakdown

Readers by professional status Count As %
Researcher 5 21%
Other 4 17%
Student > Doctoral Student 4 17%
Student > Ph. D. Student 3 13%
Student > Master 3 13%
Other 3 13%
Unknown 2 8%
Readers by discipline Count As %
Engineering 9 38%
Medicine and Dentistry 9 38%
Unspecified 1 4%
Agricultural and Biological Sciences 1 4%
Unknown 4 17%